Tech Specs
Scan Parameters
Slices: 64, with 128-slice FineRecon option
Rotation times (seconds): 0.75, 1.0, 1.5, 2.0
Data collection speed: 1,200 views per second
Minimum Slice thickness (mm): 0.625
FOV (mm): 20-500
Max scan range (in/mm): 71/1800
Max scan rotations: 100
Volume scan pitch: 0.578 to 1.578
Scan Acquisition Types
Scanogram with real-time display – AP & Lateral
Normal scan - axial mode
Volume scan - helical mode With or without Orbital Synchronization
Dynamic Scan mode for Time Density Analysis
Gantry
Gantry aperture (cm): 75
Gantry tilt: ±30 degrees
Gantry Display: Digital Panel
Scan localizer
X-ray Detector
Type: Solid state ceramic
Coverage: 40mm with 64 rows and 64-electronics channels (True64 – DAS)
Detector elements: 64 x 880 = 56,320
X-ray Tube
Anode heat capacity: 5.0MHU
Anode max. cooling rate: 748kHU/min
Tube cooling: Oil/Air
Focal Spot (mm): 0.7 x 0.8; 1.2 x 1.4
X-ray Generator
Output: 48kW
kVp selection: 80, 100, 120, 140kV
mA selection: 10-400mA (5mA steps)
Patient Table
Table-top width: 19in (475mm)
Table weight capacity: 500 lbs. (227kgs)
Horizontal accuracy: ±0.25mm
Vertical travel range: 18in (450mm) to 39in (1000mm)
Scannable range 180cm
Operator's Console
Display: single 24-inch LCD widescreen monitor
Display resolution: 1920 x 1200
Image Reconstruction
Matrix: 512 x 512
Immediate Preview Recon
Reconstruction time: up to 22 images/second (depending on recon parameter)
Iterative Reconstruction: Intelli IP Quick
CT number range: -32,768 to +32,767
Dose Reporting
Simple Dose Report
DICOM Dose Structured Report (SR)
CT Dose Check (NEMA XR-25)
MITA Smart Dose Compliant (XR-29)
Siting
Minimum suite size: 250 Sq. Ft.
Typical suite size: 340 Sq. Ft.